NDJ-5S-8S-9SO Digital Viscometer perating Instruction

Operating Instruction

-NDJ-5S-8S-9S

Digital Viscometer

CONTENT

1. Operating principle and Using …………………1

2. Main Technical Data ……………………………1

3. Application Environment ………………………2

4. Structher and Installation ………………………2

5. Using……………………………………………3

6. Panel operation…………………………………3

7. Precaution………………………………………6

8. Packing list………………………………………8  
1Operating priciple and Using

   NDJ Series Digital Viscometer (NDJ-5S-8S-9S) is a smart instrument,depending on the technology of A 16-bit high-performance single chip,which has totally changed the previous rotor method. Stepper motor runs accurately and stably as the setting speed procedure. And then torque sensor will drive rotor rotation in a constant speed. When rotor suffers viscous resistance in the tested liquid, the viscosity of the tested liquid will show on the screen by testing and handling.

   Compared with the same instrument, this one has the merit of high measure accuracy, stabled in rotate speed, wide working voltage (available in 110V, 220V, 50Hz, 60Hz) and so on. It is very convenient for users to choose the rotor and rotor velocity of the testing liquid correctly, because of the function of measured value occupies full range.

    It can widely used to measure Dynamic Viscosity of solvent based adhesive, latex, biochemicals, oil paint, coating, cosmetic, printing ink, pulp,starch, food and so on.

2: Main Technical Data

1、NDJ-1SNDJ-5S

    Measurement range10mPa.S-100,000mPa.S(1mPa.S=1cp)

    Measurement errorNDJ-1S:±5%,NDJ-5S:±3%,

    Rotor types1#, 2#, 3# and 4# rotors(0# is optional accessory)

    Rotor velocity6、12、3060RPM/min

    Dimensions95*130*155 (not including classis)

    Net weight2kg  (not including classis)

2、NDJ-8S

    Measurement range10mPa.S-2,000,000mPa.S(1mPa.S=1cp)

    Measurement error:±3%

    Rotor types1#, 2#, 3# and 4# rotors(0# is optional accessory)

    Rotor velocity0.3、 0.61.5、3、6、123060RPM/min

    Dimensions95*130*155 (not including classis)

    Net weight2kg(not including classis)

3、NDJ-9S

    Measurement range10mPa.S-6,000,000mPa.S(1mPa.S=1cp)

    Measurement error:±3%

    Rotor types1#, 2#, 3# and 4# rotors(0# is optional accessory)

    Measured Value will become larger as the rotor number grows

    Rotor velocity0.10.3、0.6、1.5、3、6、1230、60RPM/min

    Dimensions95*130*155(not including classis)

    Net weight2kg(not including classis)

3:Application Environment:

Environment temperature 5- 35

    Relative humidity:≤80%

    Power supply 220V,50Hz110V,60Hz

    There is no electro magnetic interference, severe virbation, corrosive gas around the instrument.


4:Structure and Installation

 1Structure

            As follows:

1level babble  2screen  3cover  4rotor protection frame

5the host base 6operation key 7rotor connector  

8rotor9the host base horizontal knob

2Installation

1. You need to take classis, stanchion, lifting block and hand shank from the packing box. Screwing stanchions into the hole on the base. (The rack side must face to you), tiding the hand shank of viscomeler to the lifting block.

2. Adjust the tightness of up and down by swirling the up-down button, then putT"block behind instrument into hand shank and screw them.

3. Adjust three level screws of the base to make the level bubble in the center of  black circle.

4. Take off universal-joint protective cap under the main case.

5. Connect power

5:Operation

1. Prepare the liquid to be measured and put it into a round fat container or flat container with the diameter not smaller than 60mm, you need to control temperature of measured liquid correctly (It will influence viscometer directly)

2. Please reverse screw protection bracket beneath the main case.

3. Please choose the right rotor and contrarotate in the universal joint ofthe instrument

4. Adjust the lifting screw and put the rotor into the measured liquid till the level mark on the rotor reach the liquid surface.

5. Adjust the instrument in level again.

6. During the testing, the environment temperature must be stable to make sure the show value is accurate.


6Panel operation

This instrument has eight keys: choosing rotor, setting rotor speed,,,

stat, stop,search, confirm. Turn on the power switch, enter standby state.

7Precaution

1. Because Tem is viscosity's function, so the Tem measurement error should  be  controlled in ±0.1℃  when  the  instrument works in the normal temperaure.

2. Please look out the aperture angle percentage, its value should be kept 20%-85% . When the valuce  is  too  high  or  too  low,  please change rotor or velocity. Otherwise, it will affect measurement accuracy.

3. The instrument should be used in the required range of voltage  and frequency, Othervise, it will afect measurement accuracy.

4. You should operate carefully when you put on or take off rotor.  You just need to pull up the universal joints a little bit, forbidding  pushing too  hard.

5. Afer using, universal joints and rotor should be kept clean.

6. You need to drop the instrument slowly. You'd better hold it to  avoid damage the  axis because of vibration.

7. Universal joints should be put on protctive nap when the  instrument is moved or  transported.

8. Turbid liquid, emulsion, high polymer material and some high viscosity liquid, which all  belong non-Newtonian liquid. Their apparent viscosity and velocity are not linear relationship. So,  when  you  use  different rotors, velocity, and measurement time, their viscosity measurement may be different, not the mistake of instrument testing.

9. You will get accurate result when you do as fllowing:

    A. Control the tested liquid temperature accurately.

    B. Kcep environment temperature balance.

    C. Rotor and the tested liquid must have the same temperature.

    D. The bigger rotor, higher speed usually is used for low viscometer

     liquid, and high viscometer usually uscs smaller, lower speed.

    E. Keep the surface of rotor clean.

Attach: the installation step of  0#

1): The using of 0# must be followed as instument operations. You will    know the specification by picture.

2): Pouring 25-30mL tested liquid into sleeve which has bottom.

3): Put on the out sleeve from the bottom up and fix it. (Sleeve P9  has V-groove)

4): The same way to use sleeve. When the  out  sleeve  and  rotor  are immersed in liquid, taking the out sleeve  having  V-groove  inside as the liquid line of  0#.

Packing  list

No.           Name                               Number

1      NDJ series Digital Rotary Viscometer                 1

2      Stanchion and lifting block                          1

3     Protection bracket                               1

4      Rotors, 1#, 2#, 3# and 4#                          1

5     Power line                                      1

6      Operation manual                                1

7     Quality certificate                                1

8      Guarantee card                                  1

8Attach

Installation step:

Screw 0# rotor (5); fixed tap of sleeve (2); Screw in screw (1) and fix it;

Put on size zero sleeve (4); Screw in screw (1) and fix it.



TEL:13798128437

掃碼添加微信
30分钟人性短片免费| 中文字幕乱码中文乱码51精品 | 国产99视频精品免费视频76 | 欧美性猛交xxxx富婆| 极品粉嫩嫩模大尺度无码| 国产毛多水多做爰爽爽爽| av无码免费播放| 《超大爆乳护士》在线观看| 激情97综合亚洲色婷婷五 | 入禽太深免费完整版电影| 国产成人av一区二区三区不卡| 欧美大肥婆bbbww| 少妇又紧又深又湿又爽视频| 国产JJIZZ女人多水喷水| 欧美老熟妇乱人伦人妻| 久久久亚洲欧洲日产国产成人无码| 全黄裸片一29分钟免费真人版| 亚洲AV成人无遮挡网站在线观看| 人妻含泪握住粗大| 狠狠做深爱婷婷久久综合一区| 国产一区二区三区精品| 一二三四免费bd高清视频| 日本人妻丰满熟妇久久久久久| 久久婷婷五月综合色国产香蕉 | 无码人妻精品一区二区蜜桃网站| 国产AV熟女一区二区三区 | 国产AV人人夜夜澡人人爽| 久久99精品久久久久子伦| 欧美熟妇av欧差aa片| 国产亚洲精品麻豆一区二区| 国内精品伊人久久久久AV影院| 色yeye香蕉凹凸一区二区-| 国产av精品一区二区三区| 免费人成年激情视频在线观看| 国产乱沈阳女人高潮乱叫老| 成人禁片免费播放35分钟| 久久久亚洲AV波多野结衣| 国产精品 高清 尿 小便 嘘嘘| 亚洲欧美日本韩国| 无码任你躁久久久久久老妇| 国产女人18毛片水真多1|